The ECM NOxCAN, Type G Module (NOxCANg) is a versatile and highly integratable NOx, Lambda, and O2 measurement module for the development of engines (e.g. spark-ignition engines) and combustion systems that can operate lean of stoichiometric, and their aftertreatment systems. The NOxCANg uses a ceramic NOx sensor that is mounted in the exhaust of the engine and communicates measured NOx, Lambda, O2, and all sensor parameters via its CAN port. Although designed as a measurement tool, the NOxCANg can be easily integrated into an engine or aftertreatment control strategy. The CAN node identification can be programmed by the user allowing multiple NOx modules on the same bus. Fuel H:C, O:C, and N:C ratios can be programmed. NOx sensors used with the module have memory chips in their connector where calibration information is stored. This allows the sensors to be recalibrated (zero, span) in a central location and distributed to users, ensuring consistent results throughout a large test facility. PC software to set-up, control, calibrate, and view outputs and sensor parameters is included (requires CAN adapter, not included). Two optional displays, one with programmable analog outputs, are available. These displays can be used with one or two modules.
Specifications:
Inputs:1 Ceramic NOx Sensor (Type G)
Ranges:NOx 0 to 5000 ppm (For λ > 1 only)
λ 0.4 to 25
AFR 6 to 364
%O2 0 to 25%
Pressure 0 to 517 kPa (0 to 75 psia)
Accuracies:NOx ± 5 ppm (0 to 200 ppm), ± 15 ppm (200 to 1000 ppm), ± 1.5% (elsewhere)
λ, AFR, φ ± 0.8% (at stoichiometric), ± 1.8 (average, elsewhere)
%O2 ± 0.2% (absolute)
Pressure ± 5.2 kPa (± 0.75 psia)
Response Time:Less than 1 s (NOx). Less than 150 ms (λ, AFR, φ, O2)
Fuel Type:Programmable H:C, O:C, and N:C ratios, and H2
CAN:High Speed according to ISO 11898
Configuration:Via CAN Bus with Configuration Software. Programmable Node ID.
Module:145mm x 120mm x 40mm, Environmentally Sealed
Environmental:-55 to +125°C, IP67
Sensor Cable:0.6m (sensors), +1m (standard), +2m (optional)
Power:11 to 28 VDC, AC/DC (optional)
Sensor Mounting:20mm x 1.5mm thread
Jump to Optional Kits and Parts
Related Downloads:
NOxCANg Brochure (10-08-2013, 214KB)
Aluminum Block Brochure (27KB)
Exhaust Probe Brochure (104KB)
Vboost Power Supply Brochure (165KB)
ETAS INCA Setup Guide (1052KB)
ATI VISION Setup Guide (504KB)
ECM CAN Configuration Tool (v4.33, 9591KB)
Firmware Update Procedure for Modules (225KB)
Ordering Information, Complete Kit P/N: NOxCANg
The complete kit includes the following parts:
ItemDescriptionP/NQtyPhoto
1.NOxCANg Module02-031
2.NOx Sensor (Type G)06-021
3.Lambda/NOx Cable, 1m10-021
4.Flexi-Eurofast Cable, 0.3m09-041
5.Eurofast "T" Connector09-051
6.Eurofast Termination Resistor09-061
7.Eurofast 12mm Cable, 2m09-021
8.DC Power Cable, DB9F, Banana Plugs11-021
9.20mm x 1.5mm Stainless Steel Boss and Stainless Steel Plug12-281
10.5200 Series Manuals and Configuration Software CD
This PC-based software allows configuration of CAN communication, calibration of sensors, and data logging. Software requires Kvaser (available from ECM, see below), Vector, or Peak USB-to-CAN adapter.
13-011


Optional Kits and Parts:

For pressure compensation, these additional items are included:
ItemDescriptionP/NQtyPhoto
1a.Pressure, 0-75 psia, 1/4" NPT (USA), or07-011
1b.Pressure, 0-517 kPa, 6mm (Metric)07-021
2.Pressure Cable (4-pin), 1m10-041
3a.Pressure Line Assembly, 1/4" NPT Tubing/Hose/Fittings, 28" Length (USA), or12-081
3b.Pressure Line Assembly, 6mm Tubing/Hose/Fittings, 711mm Length (Metric)12-111
4.Module Y Cable10-211

Accessories:
ItemDescriptionP/NQtyPhoto
1.CAN-to-USB Adapter13-021
2.AC/DC Power Supply w/ CAN Tap, Universal, 24V
(Requires P/N 11-17 for use with CAN-to-USB adapter)
04-011
3.Deutsch DTM3M to DB9F11-171
4.DC Power Cable, DB9F, Spades11-011
5.Vboost Power Supply (Boosts supply voltage to avoid dropout situations)04-021
6.AC/DC Power Supply, 120VAC In, 15VDC 15A Out, 4.5" x 6" 10", 13lbm
(Up to 4 Modules and 2 Display Heads)
04-041
7.AC/DC Power Supply, 120VAC In, 15VDC 30A Out, 6.75" x 6.25" x 10.25", 21lbm
(Up to 8 Modules and 4 Display Heads)
04-031
8.AC/DC Power Supply, 120VAC In, 15VDC 60A Out, 6.75" x 10.5" x 11.5", 33lbm
(Up to 16 Modules and 8 Display Heads)
04-051
9.Eurofast 12mm Cable, 10m09-03/101
10.Lambda/NOx Cable, 2m10-031
11.Pressure Cable (4-pin), 2m10-051
12.Eurofast Hub Block, 8-channel09-081
13.Termination Resistor for Hub Block09-091
14.2m Hub Power/Eurofast Harness11-101
15.1/4" UNC Module Stacking Standoff
(Use to mount and stack ECM CAN modules. One kit mounts/stacks two
modules. Each additional kit mounts/stacks an additional module.)
12-251
16.dashCAN+DASHP1
17.dashCANDASH1
18.Aluminum Mounting Plate for 8-Ch Block and Modules (16" x 10" x 0.16")12-421
19.19" Rackmount Panel (holds up to 4 display heads or 4 dashCANs)12-011
20.20mm x 1.5mm Stainless Steel Boss and Stainless Steel Plug12-281
21.1/4" NPT Mild Steel Boss and Brass Plug (USA)12-051
22.1/4" ISO Mild Steel Boss and Brass Plug (Metric)12-121
23.Aluminum Remote Sensor Mounting Block (20mm x 1.5mm, 18mm x 1.5mm, 4 x 1/4" NPT)12-311
24. Individual Cylinder Exhaust Probe (20mm x 1.5mm) (USA)12-401
25. Individual Cylinder Exhaust Probe (20mm x 1.5mm) (Metric)12-411